WEKO3
アイテム
Comparison of ICare rebound tonometer with noncontact tonometer in healthy children
https://kagawa-u.repo.nii.ac.jp/records/186
https://kagawa-u.repo.nii.ac.jp/records/18649ad42c2-9ce5-402f-8ded-ba7b740a18df
名前 / ファイル | ライセンス | アクション |
---|---|---|
![]() |
|
|
![]() |
|
Item type | 学位論文 / Thesis or Dissertation(1) | |||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
公開日 | 2013-04-17 | |||||||||||||
タイトル | ||||||||||||||
タイトル | Comparison of ICare rebound tonometer with noncontact tonometer in healthy children | |||||||||||||
言語 | en | |||||||||||||
言語 | ||||||||||||||
言語 | eng | |||||||||||||
資源タイプ | ||||||||||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_db06 | |||||||||||||
資源タイプ | doctoral thesis | |||||||||||||
著者 |
蔭山, 光代
× 蔭山, 光代
|
|||||||||||||
抄録 | ||||||||||||||
内容記述タイプ | Abstract | |||||||||||||
内容記述 | Purpose To evaluate the effectiveness of using a noncontact tonometer (NCT) versus a rebound tonometer (ICare) when measuring the intraocular pressure (IOP) in healthy children. |
|||||||||||||
言語 | en | |||||||||||||
抄録 | ||||||||||||||
内容記述タイプ | Abstract | |||||||||||||
内容記述 | Design Prospective observational study. |
|||||||||||||
言語 | en | |||||||||||||
抄録 | ||||||||||||||
内容記述タイプ | Abstract | |||||||||||||
内容記述 | Methods A total of 180 (96 males and 84 females) healthy children, aged 6 months to 15 years, were recruited. IOPs were measured using both the NCT and ICare devices in the absence of anesthesia. A successful outcome was defined as the measurement of the IOP in both eyes. The McNemar test for comparing correlated proportions was used to analyze the ICare and NCT data. |
|||||||||||||
言語 | en | |||||||||||||
抄録 | ||||||||||||||
内容記述タイプ | Abstract | |||||||||||||
内容記述 | Results Although the IOP was successfully measured in 160 children (88.9%) when using ICare, we were only able to successfully measure 130 children (72.2%) when using NCT. Below the age of 6 years, measurement of the IOP using ICare was better tolerated as compared with the NCT (McNemar test; P<0.001). The mean differences (95% limits of agreement) for the IOP readings between ICare and NCT in right and left eyes were 0.90±6.40 and 1.18±6.19 mm Hg, respectively. |
|||||||||||||
言語 | en | |||||||||||||
抄録 | ||||||||||||||
内容記述タイプ | Abstract | |||||||||||||
内容記述 | Conclusions IOP measurements performed using ICare are better tolerated in the pediatric population, as compared with measurements using NCT, especially in children below the age of 6 years. |
|||||||||||||
言語 | en | |||||||||||||
学位名 | ||||||||||||||
言語 | ja | |||||||||||||
学位名 | 博士(医学) | |||||||||||||
学位授与機関 | ||||||||||||||
識別子Scheme | kakenhi | |||||||||||||
識別子 | 16201 | |||||||||||||
言語 | ja | |||||||||||||
機関名 | 香川大学 | |||||||||||||
言語 | en | |||||||||||||
機関名 | Kagawa University | |||||||||||||
学位授与年月日 | ||||||||||||||
学位授与年月日 | 2013-04-17 | |||||||||||||
学位授与番号 | ||||||||||||||
学位授与番号 | 乙第264号 | |||||||||||||
PubMed番号 | ||||||||||||||
識別子タイプ | PMID | |||||||||||||
関連識別子 | 20179620 | |||||||||||||
関連サイト | ||||||||||||||
関連タイプ | isVersionOf | |||||||||||||
識別子タイプ | DOI | |||||||||||||
関連識別子 | https://doi.org/10.1097/IJG.0b013e3181d12dc4 | |||||||||||||
KEID | ||||||||||||||
27447 |